Pencil-beam scanning catheter for intracoronary optical coherence tomography | |
Kang, Jiqiang1; Zhu, Rui2,3,4; Sun, Yunxu1; Li, Jianan3,4; Wong, Kenneth K. Y.5,6 | |
作者部门 | 瞬态光学研究室 |
2022 | |
发表期刊 | OPTO-ELECTRONIC ADVANCES
![]() |
ISSN | 2096-4579 |
卷号 | 5期号:3 |
产权排序 | 4 |
摘要 | Current gradient-index (GRIN) lens based proximal-driven intracoronary optical coherence tomography (ICOCT) probes consist of a spacer and a GRIN lens with large gradient constant. This design provides great flexibility to control beam profiles, but the spacer length should be well controlled to obtain desired beam profiles and thus it sets an obstacle in mass catheter fabrication. Besides, although GRIN lens with large gradient constant can provide tight focus spot, it has short depth of focus and fast-expanded beam which leads to poor lateral resolution for deep tissue. In this paper, a type of spacer-removed probe is demonstrated with a small gradient constant GRIN lens. This design simplifies the fabrication process and is suitable for mass production. The output beam of the catheter is a narrow nearly collimated light beam, referred to as pencil beam here. The full width at half maximum beam size varies from 35.1 mu m to 75.3 mu m in air over 3-mm range. Probe design principles are elaborated with probe/catheter fabrication and performance test. The in vivo imaging of the catheter was verified by a clinical ICOCT system. Those results prove that this novel pencil-beam scanning catheter is potentially a good choice for ICOCT systems. |
关键词 | optical coherence tomography endoscopic imaging intravascular imaging fiber optics imaging |
DOI | 10.29026/oea.2022.200050 |
收录类别 | SCI ; EI |
语种 | 英语 |
WOS记录号 | WOS:000778790100002 |
出版者 | CAS, INST OPTICS & ELECTRONICS, ED OFF OPTO-ELECTRONIC JOURNALS |
EI入藏号 | 20221511951912 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.opt.ac.cn/handle/181661/95817 |
专题 | 瞬态光学研究室 |
通讯作者 | Sun, Yunxu; Wong, Kenneth K. Y. |
作者单位 | 1.Harbin Inst Technol Shenzhen, Sch Elect & Informat Engn, Shenzhen 518055, Peoples R China 2.Tsinghua Univ, Int Grad Sch Shenzhen, Shenzhen 518055, Peoples R China 3.Shenzhen Vivolight Med Device & Technol Co Ltd, Shenzhen 518055, Peoples R China 4.Chinese Acad Sci, Xian Inst Opt & Precis Mech, State Key Lab Transient Opt & Photon, Xian 710119, Peoples R China 5.Univ Hong Kong, Dept Elect & Elect Engn, Pokfulam Rd, Hong Kong, Peoples R China 6.Adv Biomed Instrumentat Ctr, Shatin, Hong Kong Sci Pk, Hong Kong, Peoples R China |
推荐引用方式 GB/T 7714 | Kang, Jiqiang,Zhu, Rui,Sun, Yunxu,et al. Pencil-beam scanning catheter for intracoronary optical coherence tomography[J]. OPTO-ELECTRONIC ADVANCES,2022,5(3). |
APA | Kang, Jiqiang,Zhu, Rui,Sun, Yunxu,Li, Jianan,&Wong, Kenneth K. Y..(2022).Pencil-beam scanning catheter for intracoronary optical coherence tomography.OPTO-ELECTRONIC ADVANCES,5(3). |
MLA | Kang, Jiqiang,et al."Pencil-beam scanning catheter for intracoronary optical coherence tomography".OPTO-ELECTRONIC ADVANCES 5.3(2022). |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Pencil-beam scanning(1926KB) | 期刊论文 | 出版稿 | 限制开放 | CC BY-NC-SA | 请求全文 |