The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging | |
Li, Yi1,2,4; Liu, HongJun1,3![]() ![]() ![]() | |
作者部门 | 瞬态光学研究室 |
2022-02 | |
发表期刊 | TECHNOLOGY IN CANCER RESEARCH & TREATMENT
![]() |
ISSN | 1533-0346;1533-0338 |
卷号 | 21 |
产权排序 | 1 |
摘要 | Purpose: The aim of this study is to measure the output factor (OF) and profile of surface dose in regular and small radiation therapy fields using Cherenkov imaging (CI). Methods: A medical linear accelerator (linac) was employed to generate radiation fields, including regular open photon field (ROPF), regular wedge photon field (RWPF), regular electron field (REF) and small photon field (SPF). The photon beams consisted of two filter modes including flattening filter (FF) and flattening filter free (FFF). All fields were delivered to a solid water phantom. Cherenkov light was captured using a charge-coupled device system during phantom irradiation. The OF and profile of surface dose measured by CI were compared with those determined by film measurement, ionization chamber measurement and treatment planning system calculation in order to examine the feasibility of measuring surface dose OF and profile using CI. Results: The discrepancy between surface dose OF measured by CI and that determined by other methods is less than 6% in ROPFs with size less than 10 x 10 cm(2), REFs with size less than 10 x 10 cm(2), and SPFs except for 1 x 1 cm(2) field. In the flat profile region, the discrepancy between surface dose profile measured by CI and that determined by other methods is less than 4% in REFs and less than 3% in ROPFs, RWPFs, and SPFs except for 1 x 1 cm2 field. The discrepancy of the surface dose profile is in compliance with the recommendation by IAEA TRS 430 reports. The discrepancy between field width measured by CI and that determined by film measurement is equal to or less than 2 mm, which is within the tolerance recommend by the guidelines of linac quality assurance in regular open FF photon fields, SPFs, and REFs with cone size of 10 x 10 cm(2) in area. Conclusion: CI can be used to quantitatively measure the OF and profile of surface dose. It is feasible to use CI to measure the surface dose profile and field width in regular open FF photon fields and SPFs except for 1 x 1 cm(2) field. |
关键词 | Radiotherapy quality assurance Cherenkov imaging surface dose output factor profile |
DOI | 10.1177/15330338211073432 |
收录类别 | SCI |
语种 | 英语 |
WOS记录号 | WOS:000758230700001 |
出版者 | SAGE PUBLICATIONS INC |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.opt.ac.cn/handle/181661/95738 |
专题 | 瞬态光学研究室 |
通讯作者 | Liu, HongJun; Zhang, Chunmin |
作者单位 | 1.Chinese Acad Sci, Xian Inst Opt & Precis Mech, State Key Lab Transient Opt & Photon, Xian 710119, Peoples R China 2.Xi An Jiao Tong Univ, Sch Phys, Xian 710049, Peoples R China 3.Shanxi Univ, Collaborat Innovat Ctr Extreme Opt, Taiyuan 030006, Peoples R China 4.Univ Chinese Acad Sci, Beijing 100049, Peoples R China |
推荐引用方式 GB/T 7714 | Li, Yi,Liu, HongJun,Huang, Nan,et al. The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging[J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT,2022,21. |
APA | Li, Yi,Liu, HongJun,Huang, Nan,Wang, Zhaolu,&Zhang, Chunmin.(2022).The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging.TECHNOLOGY IN CANCER RESEARCH & TREATMENT,21. |
MLA | Li, Yi,et al."The Measurement of the Surface Dose in Regular and Small Radiation Therapy Fields Using Cherenkov Imaging".TECHNOLOGY IN CANCER RESEARCH & TREATMENT 21(2022). |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
The Measurement of t(11107KB) | 期刊论文 | 出版稿 | 限制开放 | CC BY-NC-SA | 请求全文 |