OPT OpenIR  > 检测技术研究中心
Design and Calibration of Weak-Light Single Star Simulator
其他题名弱光单星模拟器的设计与标定
Liu, Shangkuo1,2; Xue, Xun1; Li, Kun1; Cao, Kun1; Zhao, Jianke1; Zhou, Yan1; Yao, Baoli1
作者部门检测技术研究中心
2017-10-10
发表期刊Guangxue Xuebao/Acta Optica Sinica
ISSN02532239
卷号37期号:10
产权排序1
摘要

Aiming at the problem of high magnitude target simulation in the lab, a weak-light single star simulator (WLSSS) consisting of light source, adjustable diaphragm, integrating sphere, photoelectric detector and collimator is developed to measure the magnitude detecting ability of detecting cameras and star sensors. The relationship between the readings of photoelectric detector, the spectral radiance in the outlet of integrating sphere and tunable obscuration ratio is deduced. Combining the magnitude definition formula and the image illuminance equation formula, the operation principle of weak-light single star simulator is introduced, and the calibration challenge of high magnitude target is solved. The magnitude simulation range and precision of WLSSS are theoretically analyzed, and results show that the highest simulating magnitude is 19.5 Mv, and the precision is 11.6%. When the simulation magnitude is lower than 15 Mv, its precision is better than 8%. Experiment results show that in the lab, the maximum relative error between simulated magnitudes and KLL-04 wide-range illuminator tested magnitudes is 7.09%. The relative error of measured magnitude detecting ability between laboratory and astro-observation detecting cameras is 1.9% in 6.5 Mv, and 2.6% in 15.2 Mv. The designed WLSSS can effectively simulate high magnitude targets. © 2017, Chinese Lasers Press. All right reserved.

DOI10.3788/AOS201737.1012001
收录类别EI ; CSCD
语种中文
引用统计
文献类型期刊论文
条目标识符http://ir.opt.ac.cn/handle/181661/29449
专题检测技术研究中心
作者单位1.Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an; Shaanxi; 710119, China
2.University of Chinese Academy of Sciences, Beijing; 100049, China
推荐引用方式
GB/T 7714
Liu, Shangkuo,Xue, Xun,Li, Kun,et al. Design and Calibration of Weak-Light Single Star Simulator[J]. Guangxue Xuebao/Acta Optica Sinica,2017,37(10).
APA Liu, Shangkuo.,Xue, Xun.,Li, Kun.,Cao, Kun.,Zhao, Jianke.,...&Yao, Baoli.(2017).Design and Calibration of Weak-Light Single Star Simulator.Guangxue Xuebao/Acta Optica Sinica,37(10).
MLA Liu, Shangkuo,et al."Design and Calibration of Weak-Light Single Star Simulator".Guangxue Xuebao/Acta Optica Sinica 37.10(2017).
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Design and Calibrati(1647KB)期刊论文出版稿限制开放CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Liu, Shangkuo]的文章
[Xue, Xun]的文章
[Li, Kun]的文章
百度学术
百度学术中相似的文章
[Liu, Shangkuo]的文章
[Xue, Xun]的文章
[Li, Kun]的文章
必应学术
必应学术中相似的文章
[Liu, Shangkuo]的文章
[Xue, Xun]的文章
[Li, Kun]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。