KMS Xi'an Institute of Optics and Precision Mechanics,CAS
Simulation of the impact of using a novel neutron conversion screen on detector time characteristics and efficiency | |
Liu, Yiheng1,2,3; He, Kai1![]() ![]() ![]() | |
Department | 条纹相机工程中心 |
2022-04-01 | |
Source Publication | AIP Advances
![]() |
ISSN | 21583226 |
Volume | 12Issue:4 |
Contribution Rank | 1 |
Abstract | To directly measure the DT neutrons from inertial confinement fusion with a high time resolution, a new type of neutron conversion composed of a CH2 conversion layer, a metal moderation layer, and a CsI secondary electron emission layer is proposed. The conversion screen is based on the principle that recoil protons produced by elastic scattering of the neutrons in CH2 interact with CsI to generate secondary electrons. The moderation layer can filter the energy spectrum of protons to prevent low-energy protons from reaching CsI, which shortens the duration of the secondary electron pulse and improves the temporal resolution of the conversion screen. Based on the Monte Carlo method, both the neutron impulse and background γ-rays response of this conversion screen were calculated. The simulation indicates that the temporal resolution of the conversion screen can reach up to 4.9 ps when the thickness of the gold layer is 100 μm. The detection efficiency of secondary electrons/neutrons can reach 7.4 × 10-3. The detection efficiency of the neutron conversion screen for secondary electrons/γ-rays is an order of magnitude lower than the neutron impulse response, and the response time of γ-rays is 20 ps earlier than the neutron pulses. This means that using this conversion screen is beneficial to distinguish between neutrons and γ-rays and has a good signal-to-noise ratio. © 2022 Author(s). |
DOI | 10.1063/5.0073025 |
Indexed By | SCI ; EI |
Language | 英语 |
WOS ID | WOS:000792000400002 |
Publisher | American Institute of Physics Inc. |
EI Accession Number | 20221712039946 |
Citation statistics | |
Document Type | 期刊论文 |
Identifier | http://ir.opt.ac.cn/handle/181661/95857 |
Collection | 条纹相机工程中心 |
Corresponding Author | He, Kai; Tian, Jinshou |
Affiliation | 1.Key Laboratory of Ultra-fast Photoelectric Diagnostics Technology, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an; 710119, China; 2.University of Chinese Academy of Sciences, Beijing; 100049, China; 3.Collaborative Innovation Center of Extreme Optics, Shanxi University, Taiyuan; 030006, China; 4.The Northwest Institute of Nuclear Technology, Xi'an; 710024, China; 5.Key Laboratory for Laser Plasmas (MoE), Department of Physics and Astronomy, Shanghai Jiao Tong University, Shanghai; 200240, China |
Recommended Citation GB/T 7714 | Liu, Yiheng,He, Kai,Wang, Gang,et al. Simulation of the impact of using a novel neutron conversion screen on detector time characteristics and efficiency[J]. AIP Advances,2022,12(4). |
APA | Liu, Yiheng.,He, Kai.,Wang, Gang.,Gao, Guilong.,Yan, Xin.,...&Tian, Jinshou.(2022).Simulation of the impact of using a novel neutron conversion screen on detector time characteristics and efficiency.AIP Advances,12(4). |
MLA | Liu, Yiheng,et al."Simulation of the impact of using a novel neutron conversion screen on detector time characteristics and efficiency".AIP Advances 12.4(2022). |
Files in This Item: | ||||||
File Name/Size | DocType | Version | Access | License | ||
Simulation of the im(6893KB) | 期刊论文 | 出版稿 | 开放获取 | CC BY-NC-SA | View Application Full Text |
Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.