OPT OpenIR  > 检测技术研究中心
Calibration of stray light based on point source transmittance measurement system
其他题名基于点源透过率测试系统的杂散光标定
Xu, Liang1,2; Gao, Li-Min1; Zhao, Jian-Ke1; Liu, Feng1; Zhou, Yan1; Li, Zhao-Hui1,2; Yang, Fei3; Zhao, Qing1,2; Xu, Liang (xuliang757@163.com)
作者部门检测技术研究中心
2016-07-01
发表期刊Guangxue Jingmi Gongcheng/Optics and Precision Engineering
ISSN1004924X
卷号24期号:7页码:1607-1614
产权排序1
摘要To improve the stray light testing ability and calibration accuracy of the stray light equipment used in test of the Point Source Transmittance (PST), a calibration lens was proposed to calibrate the testing range and testing accuracy of the equipment in a large off;axial angle. By using a simple physical model, the calibration lens was designed in a laboratory. The physical parameters of the calibration lens were measured, then these parameters were taken into the TracePro to calculate the PSTs in different off;axial angles. Finally, the calculated result and the tested result of PSTs were compared, and the test accuracy of the equipment were obtained. The experimental results indicate that the difference between calculate results and testing results of the calibration lens is better than lg/0.5, meeting the need of testing accuracy calibration of the stray light equipment in testing PSTs and providing a reliable reference for the absolute measurement of PSTs. This technology solves the problem in testing accuracy calibration of PST measuring systems. © 2016, Science Press. All right reserved.
关键词Ability Testing Distribution Functions Equipment Stray Light
DOI10.3788/OPE.20162407.1607
收录类别EI
语种中文
引用统计
文献类型期刊论文
条目标识符http://ir.opt.ac.cn/handle/181661/28256
专题检测技术研究中心
通讯作者Xu, Liang (xuliang757@163.com)
作者单位1.Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an; 710119, China
2.University of Chinese Academy of Sciences, Beijing; 100049, China
3.AVIC Xi'an Aircraft Industry (Group) Company LTD., Xi'an; 710089, China
推荐引用方式
GB/T 7714
Xu, Liang,Gao, Li-Min,Zhao, Jian-Ke,et al. Calibration of stray light based on point source transmittance measurement system[J]. Guangxue Jingmi Gongcheng/Optics and Precision Engineering,2016,24(7):1607-1614.
APA Xu, Liang.,Gao, Li-Min.,Zhao, Jian-Ke.,Liu, Feng.,Zhou, Yan.,...&Xu, Liang .(2016).Calibration of stray light based on point source transmittance measurement system.Guangxue Jingmi Gongcheng/Optics and Precision Engineering,24(7),1607-1614.
MLA Xu, Liang,et al."Calibration of stray light based on point source transmittance measurement system".Guangxue Jingmi Gongcheng/Optics and Precision Engineering 24.7(2016):1607-1614.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Calibration of stray(659KB)期刊论文作者接受稿限制开放CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Xu, Liang]的文章
[Gao, Li-Min]的文章
[Zhao, Jian-Ke]的文章
百度学术
百度学术中相似的文章
[Xu, Liang]的文章
[Gao, Li-Min]的文章
[Zhao, Jian-Ke]的文章
必应学术
必应学术中相似的文章
[Xu, Liang]的文章
[Gao, Li-Min]的文章
[Zhao, Jian-Ke]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。