Cooled Dyson long-wave infrared push-broom imaging spectrometer by re-imaging | |
Sun, Jiayin; Liu, Ying; Jiang, Yang; Li, Chun; Sun, Qiang; Hu, Xinrong; Sun, JY | |
作者部门 | 研究生部 |
2016-05-15 | |
发表期刊 | OPTICS COMMUNICATIONS
![]() |
ISSN | 0030-4018 |
卷号 | 367页码:274-278 |
产权排序 | 3 |
摘要 | A cooled long-wave infrared push-broom imaging spectrometer with an F-number of 2 was designed based on the Dyson configuration. A three-mirror off-axis aspherical optical system that provided excellent slit-shaped images was selected as the fore telescope objective. The re-imaging method was applied to obtain a cold stop efficiency of 100%, and the corrector lens in traditional Dyson imaging spectrometers was replaced with re-imaging lenses to correct spherical aberrations. The designed imaging spectrometer provided a spectral resolution of 25 nm at a range of 8-12 mu m and possessed a relatively small volume. (C) 2016 Elsevier B.V. All rights reserved. |
关键词 | Long-wave Infrared Imaging Spectrometer Dyson Configuration Re-imaging |
学科领域 | Optics |
DOI | 10.1016/j.optcom.2016.01.067 |
收录类别 | SCI ; EI |
语种 | 英语 |
WOS研究方向 | Optics |
WOS记录号 | WOS:000370127300042 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.opt.ac.cn/handle/181661/27827 |
专题 | 研究生部 |
通讯作者 | Sun, JY |
推荐引用方式 GB/T 7714 | Sun, Jiayin,Liu, Ying,Jiang, Yang,et al. Cooled Dyson long-wave infrared push-broom imaging spectrometer by re-imaging[J]. OPTICS COMMUNICATIONS,2016,367:274-278. |
APA | Sun, Jiayin.,Liu, Ying.,Jiang, Yang.,Li, Chun.,Sun, Qiang.,...&Sun, JY.(2016).Cooled Dyson long-wave infrared push-broom imaging spectrometer by re-imaging.OPTICS COMMUNICATIONS,367,274-278. |
MLA | Sun, Jiayin,et al."Cooled Dyson long-wave infrared push-broom imaging spectrometer by re-imaging".OPTICS COMMUNICATIONS 367(2016):274-278. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Cooled Dyson long-wa(1033KB) | 期刊论文 | 作者接受稿 | 限制开放 | CC BY-NC-SA | 请求全文 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论