A Full Aperture Backscattering Light Diagnostic System Installed on the Shenguang-III Prototype Laser Facility | |
Xu Tao1; Mei Yu1; Wei Huiyue1; Peng Xiaoshi1; Wang Feng1; Yang Dong1; Liu Shenye1; Yan Yadong2![]() | |
作者部门 | 先进光学仪器研究室 |
2014-06-01 | |
发表期刊 | PLASMA SCIENCE & TECHNOLOGY
![]() |
ISSN | 1009-0630 |
卷号 | 16期号:6页码:567-570 |
产权排序 | 2 |
摘要 | A full aperture backscattering light diagnostic system (FABLDS) implemented on the Shen Guang-III Prototype Laser Facility is described in the paper. FABLDS measures both stimulated brillouin scattering (SBS) and stimulated Raman scattering (SRS) with a series of optical detectors. Energy sensors record the integrated energy, and streak cameras coupled with spectrometers measure the temporal spectrum of the backscattering light. This paper provides an overview of the FABLDS and detailed descriptions of the optical path. Special components, including off-axis parabolic mirror, spatial filter and optical light filters, are incorporated along the beam path for purifying the scattering light. Several hohlraum targets were employed, including C5II12 gas-filled targets and empty targets in the experiments. Results presented in the paper indicate that the fraction of backscatter light has been obviously shrinked when the laser is smoothed by continuous phase plates (CPP). |
文章类型 | Article |
关键词 | Backscattering Light Icf Off-axis Parabolic Mirror Hohlraum Targets Cpp |
WOS标题词 | Science & Technology ; Physical Sciences |
DOI | 10.1088/1009-0630/16/6/05 |
收录类别 | SCI ; EI |
关键词[WOS] | NATIONAL-IGNITION-FACILITY ; PLASMA INTERACTIONS |
语种 | 英语 |
WOS研究方向 | Physics |
WOS类目 | Physics, Fluids & Plasmas |
WOS记录号 | WOS:000337678400005 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.opt.ac.cn/handle/181661/22484 |
专题 | 先进光学仪器研究室 |
作者单位 | 1.China Acad Engn Phys, Res Ctr Laser Fus, Mianyang 621900, Peoples R China 2.Chinese Acad Sci, Xian Inst Opt & Precis Mech, Xian 721000, Peoples R China |
推荐引用方式 GB/T 7714 | Xu Tao,Mei Yu,Wei Huiyue,et al. A Full Aperture Backscattering Light Diagnostic System Installed on the Shenguang-III Prototype Laser Facility[J]. PLASMA SCIENCE & TECHNOLOGY,2014,16(6):567-570. |
APA | Xu Tao.,Mei Yu.,Wei Huiyue.,Peng Xiaoshi.,Wang Feng.,...&Yan Yadong.(2014).A Full Aperture Backscattering Light Diagnostic System Installed on the Shenguang-III Prototype Laser Facility.PLASMA SCIENCE & TECHNOLOGY,16(6),567-570. |
MLA | Xu Tao,et al."A Full Aperture Backscattering Light Diagnostic System Installed on the Shenguang-III Prototype Laser Facility".PLASMA SCIENCE & TECHNOLOGY 16.6(2014):567-570. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
A Full Aperture Back(379KB) | 期刊论文 | 出版稿 | 限制开放 | CC BY | 请求全文 |