Study of Ultraviolet Photon Counting Detector with Delay-Line Anode | |
其他题名 | 延迟线阳极紫外光子探测器研究 |
Lei, Fan-Pu1,2,3; Bai, Yong-Lin3; Zhu, Bing-Li3; Bai, Xiao-Hong3; Qin, Jun-Jun3; Xu, Peng3; Hou, Xun1 | |
作者部门 | 精密物理量测量实验室 |
2017-10-01 | |
发表期刊 | Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis |
ISSN | 10000593 |
卷号 | 37期号:10页码:2989-2994 |
产权排序 | 1 |
摘要 | Delay-line anode detector based on MCP electron multiplier detect the incident photon's location by the arrival time difference of the MCP output charge pulse at each end of delay line. Due to their high spatial resolution and high counting rate, photon counting detectors with delay line anode have been widely used in ultraviolet spectral imaging systems. Operational principle of ultraviolet photon detector based on delay line anode is analyzed, and a novel two-dimensional serpentine delay line anode is designed. Electron cloud output from MCP is collected only by the anode surface. The upper serpentine delay line receives the MCP output charge directly, while the lower serpentine delay line receives the charge using a series of pads and vias. This anode is made of printed circuit board instead of laser etching, which greatly simplifies the fabrication process. Photon counting detector with this type of anode achieves a 92 μm FWHM resolution. And the image linearity is better than 100 μm. The results of theory analysis and experiments demonstrated the feasibility of ultraviolet photon imaging system. It provides a theoretical basis and experiments guidance for the development of ultraviolet imaging spectrometers. © 2017, Peking University Press. All right reserved. |
DOI | 10.3964/j.issn.1000-0593(2017)10-2989-06 |
收录类别 | SCI ; EI ; CSCD |
语种 | 中文 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.opt.ac.cn/handle/181661/29986 |
专题 | 条纹相机工程中心 |
作者单位 | 1.State Key Laboratory of Transient Optics and Photonics, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an, 710119, China 2.University of Chinese Academy of Sciences, Beijing, 100049, China 3.Key Laboratory of Ultrafast Diagnostic Technique, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an, 710119, China |
推荐引用方式 GB/T 7714 | Lei, Fan-Pu,Bai, Yong-Lin,Zhu, Bing-Li,et al. Study of Ultraviolet Photon Counting Detector with Delay-Line Anode[J]. Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis,2017,37(10):2989-2994. |
APA | Lei, Fan-Pu.,Bai, Yong-Lin.,Zhu, Bing-Li.,Bai, Xiao-Hong.,Qin, Jun-Jun.,...&Hou, Xun.(2017).Study of Ultraviolet Photon Counting Detector with Delay-Line Anode.Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis,37(10),2989-2994. |
MLA | Lei, Fan-Pu,et al."Study of Ultraviolet Photon Counting Detector with Delay-Line Anode".Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis 37.10(2017):2989-2994. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Study of Ultraviolet(1588KB) | 期刊论文 | 作者接受稿 | 开放获取 | CC BY-NC-SA | 浏览 请求全文 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论